Details : C10RMS20 - creatine kinase (CK) in Calibration solution, Blood serum
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C10RMS20 |
| Reference Laboratory Status | EQAS Organizer |
| Analyte Name | creatine kinase (CK) |
| IUPAC IFCC Number | EC 2.7.3.2 |
| Applicable Matrices | matrix of calibrators and control materials |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | kinetic spectrophotometry for enzyme activities |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | IFCC reference measurement procedure (37 °C) for CK |
| JCTLM Database Identifier Number | NRMeth 65 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Catalytic activity concentration |
| From (0) | 1 |
| To (0) | 25 |
| Unit (0) | µkat/L |
| From (1) | 5 |
| To (1) | 1.6 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Survey Organisation | JCTLM, IFCC, RfB, RELA |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 18. |
| Organization Acronym | NCCL |
| Organization Name | National Center for Clinical Laboratories |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Enzymes |