JCTLM Database: higher-order reference materials, methods and services
Details : C10RMS22 - uric acid in Calibration solution, Blood serum
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C10RMS22 |
| Reference Laboratory Status | EQAS Organizer |
| Analyte Name | uric acid |
| IUPAC IFCC Number | CAS69-93-2 |
| Applicable Matrices | matrix of calibrators and control materials |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | ID-LC/MS/MS |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | NIM reference method
for serum uric acid |
| JCTLM Database Identifier Number | C7RMP27 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount-of substance
concentration |
| From (0) | 90 |
| To (0) | 900 |
| Unit (0) | µmol/L |
| From (1) | 2.5 |
| To (1) | 2.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Reference Procedure Status Acceptable | Yes |
| Laboratory Status | (b) Accredited (ISO 17025+15195) |
| Survey Organisation | JCTLM, IFCC, RfB, RELA |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 18. |
| Participation Status in Intercomparison Surveys acceptable | Yes |
| Comments of Review Teams | Recommended for Listing |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | NCCL |
| Organization Name | National Center for Clinical Laboratories |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Metabolites and substrates |