Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C10RMS22 - uric acid in Calibration solution, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC10RMS22
Reference Laboratory StatusEQAS Organizer
Analyte Nameuric acid
IUPAC IFCC NumberCAS69-93-2
Applicable Matricesmatrix of calibrators and control materials
Traceable to SIYes
Principle of Analytical ProcedureID-LC/MS/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureNIM reference method for serum uric acid
JCTLM Database Identifier NumberC7RMP27
Procedure Difference in Your LaboratoryNo
QuantityAmount-of substance concentration
From (0)90
To (0)900
Unit (0)µmol/L
From (1)2.5
To (1)2.5
Unit (1)%
Level of Confidence (%)95%
Reference Procedure Status AcceptableYes
Laboratory Status(b) Accredited (ISO 17025+15195)
Survey OrganisationJCTLM, IFCC, RfB, RELA
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 18.
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsRecommended for Listing
Recommended for Publication in the JCTLM ListYes
Organization AcronymNCCL
Organization NameNational Center for Clinical Laboratories
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates