Details : C10RMS23 - creatinine in Calibration solution, Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C10RMS23 |
Reference Laboratory Status | EQAS Organizer |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Applicable Matrices | matrix of calibrators and control materials |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | Isotope dilution of creatinine by LCMS, Org-022 |
JCTLM Database Identifier Number | CYC_II_L_I_RMP_07 |
Procedure Difference in Your Laboratory | Yes |
Quantity | Amount-of substance
concentration |
From (0) | 40 |
To (0) | 1500 |
Unit (0) | µmol/L |
From (1) | 1.5 |
To (1) | 1.5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | (b) Accredited (ISO 17025+15195) |
Survey Organisation | JCTLM, IFCC, RfB, RELA |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 18. |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing |
Recommended for Publication in the JCTLM List | Yes |
Validation Report for Improved Method | Validation_Report_C10RMS23.pdf |
Organization Acronym | NCCL |
Organization Name | National Center for Clinical Laboratories |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |