Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C10RMS9 - total bilirubin in High purity material, Calibration solution, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC10RMS9
Reference Laboratory StatusHospital of University
Analyte Nametotal bilirubin
IUPAC IFCC NumberCAS635-65-4
Traceable to SIYes
Principle of Analytical ProcedureSpectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureDoumas reference method for total bilirubin
JCTLM Database Identifier NumberNRMeth 29
Procedure Difference in Your LaboratoryNo
QuantityAmount-of substance concentration
From (0)37.21
To (0)255.77
Unit (0)µmol/L
From (1)1
To (1)0.5
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from 37.31 µmol/L to 255.77 µmol/L without dilution of the sample
Comments on Uncertainty RangeThe expanded uncertainty calculated according to GUM and was the minimum expanded uncertainty obtained from RELA survey.
Reference Procedure Status AcceptableYes
Laboratory Status(b) Accredited (ISO 17025+15195)
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 65.
Additional Information 1No. 65 is the number of our lab participation in RELA - IFCC External Quality assessment scheme for Reference Laboratories.
Participation Status in Intercomparison Surveys acceptableYes
RemarksIncluding PT measurement provided by China National Center for Clinical Laboratories and RELA - IFCC External Quality assessment scheme for Reference Laboratories, and lab assigned value for Guangdong Center for Clinical Laboratories, Sichuan Maker Biotechnology CO. LTD, Shenzhen Mindray Bio-medical Electronics CO. LTD, and China National Accreditation Service for Conformity Assessment.
Comments of Review TeamsRecommended for Listing
Recommended for Publication in the JCTLM ListYes
Organization AcronymGPHCM
Organization NameGuangdong Provincial Hospital of Chinese Medicine
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates