Details : C11RMP4 - INSTAND Reference measurement procedure for lithium
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C11RMP4 |
| Analyte Name | lithium |
| IUPAC IFCC Number | CAS7439-93-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen blood serum, or blood plasma |
| Name | INSTAND Reference measurement procedure for lithium |
| Organization that developed Validated Method | INSTAND e. V. |
| Measurement Techniques Used | Inductively Coupled Plasma-(Isotope Dilution) Sector Field Mass Spectrometry (ICP-(ID) SFMS) |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.1 |
| To (0) | 5 |
| Unit (0) | mmol/l |
| From (1) | 1 |
| To (1) | 1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Clin. Lab., 2013, 59, 1017-1029
|
| Comments to be published via JCTLM DB | The reference measurement procedure gives total concentration of the respective analyte and not the ionized form. |
| Review Team Name | electrolytes |
| Analyte Category Name | Electrolytes |
| Measurement principle/technique | Inductively coupled plasma mass spectrometry |