JCTLM Database: higher-order reference materials, methods and services
Details : C11RMP5 - INSTAND Reference measurement procedure for magnesium
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C11RMP5 |
Analyte Name | magnesium |
IUPAC IFCC Number | CAS7439-95-4 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen blood serum, or blood plasma |
Name | INSTAND Reference measurement procedure for magnesium |
Organization that developed Validated Method | INSTAND e. V. |
Measurement Techniques Used | Inductively Coupled Plasma-(Isotope Dilution) Sector Field Mass Spectrometry (ICP-(ID) SFMS) |
Quantity | Amount-of-substance concentration |
From (0) | 0.1 |
To (0) | 5 |
Unit (0) | mmol/l |
From (1) | 1 |
To (1) | 1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Clin. Lab., 2013, 59, 1017-1029
|
Comments to be published via JCTLM DB | The reference measurement procedure gives total concentration of the respective analyte and not the ionized form. |
Review Team Name | electrolytes |
Analyte Category Name | Electrolytes |
Measurement principle/technique | Inductively coupled plasma mass spectrometry |