Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C11RMP7 - INSTAND Reference measurement procedure for sodium

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC11RMP7
Analyte Namesodium
IUPAC IFCC NumberCAS7440-23-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen blood serum, or blood plasma
NameINSTAND Reference measurement procedure for sodium
Organization that developed Validated MethodINSTAND e. V.
Measurement Techniques UsedInductively Coupled Plasma-(Isotope Dilution) Sector Field Mass Spectrometry (ICP-(ID) SFMS)
QuantityAmount-of-substance concentration
From (0)70
To (0)200
Unit (0)mmol/l
From (1)1
To (1)1
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationClin. Lab., 2013, 59, 1017-1029
Comments to be published via JCTLM DBThe reference measurement procedure gives total concentration of the respective analyte and not the ionized form. Results have been obtained for samples showing protein concentration between 45.4 g/L and 87.4 g/L. The unit is mmol/L of serum including lipids and proteins.
Review Team Nameelectrolytes
Analyte Category NameElectrolytes
Measurement principle/techniqueInductively coupled plasma mass spectrometry