JCTLM Database: higher-order reference materials, methods and services
Details : C11RMS1 - cortisol in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C11RMS1 |
Reference Laboratory Status | University |
Analyte Name | cortisol |
IUPAC IFCC Number | CAS50-23-7 |
Applicable Matrices | lyophilized, fresh, or frozen serum |
Traceable to SI | yes |
Principle of Analytical Procedure | ID/LC/MS/MS |
Reference Method Listed in JCTLM | based on JCTLM listed NRMeth 8; NRMeth 95 and NRMeth 38 |
Name of JCTLM Listed Reference Method Procedure | NRMeth 8; NRMeth 95 and NRMeth 38 |
JCTLM Database Identifier Number | NRMeth 8 |
Procedure Difference in Your Laboratory | Yes |
Quantity | Amount-of-substance concentration |
From (0) | 65 |
To (0) | 2000 |
Unit (0) | nmol/L |
From (1) | 1.5 |
To (1) | 3.3 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is calculated for measurement protocol n = 6 |
Laboratory Status | B |
Survey Organisation | IFCC RELA |
Intercomparison Results Publicly Available | yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Recommended for Publication in the JCTLM List | Yes |
Validation Report for Improved Method | Method_Validation_Report_C11RMS1.pdf |
Organization Acronym | UGent - Ref4U |
Organization Name | Laboratory of Toxicology, Faculty of Pharmaceutical Sciences, Ghent University |
Organization Website | https://www.ugent.be/fw/bioanalysis/en/research/toxicology/services/ref4u.htm |
Organization Country ISO Code | BE |
Organization Country Name | Belgium |
Analyte Category Name | Non-peptide hormones |