Details : C11RMS11 - 17β-estradiol in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC11RMS11
Reference Laboratory StatusEQAS Organizer
Analyte Name17β-estradiol
IUPAC IFCC NumberCAS50-28-2
Applicable Matriceslyophilized, fresh, or frozen blood serum or plasma
Traceable to SIYes
Principle of Analytical ProcedureGC-ID/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureUniversity of Ghent reference measurement procedure for estradiol
JCTLM Database Identifier NumberNRMeth 10
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)37
To (0)2500
Unit (0)pmol/L
From (1)1
To (1)1
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe stated expanded uncertainty value corresponds to the best measurement capability.
Survey OrganisationIFCC RELA Surveys
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Recommended for Publication in the JCTLM ListYes
Organization AcronymINSTAND e. V.
Organization NameSociety for Promoting Quality Assurance in Medical Laboratories
Organization Websitehttp://www.instand-ev.de
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameNon-peptide hormones