Details : C11RMS11 - 17β-estradiol in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C11RMS11 |
Reference Laboratory Status | EQAS Organizer |
Analyte Name | 17β-estradiol |
IUPAC IFCC Number | CAS50-28-2 |
Applicable Matrices | lyophilized, fresh, or frozen blood serum or plasma |
Traceable to SI | Yes |
Principle of Analytical Procedure | GC-ID/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | University of Ghent reference measurement procedure for estradiol |
JCTLM Database Identifier Number | NRMeth 10 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 37 |
To (0) | 2500 |
Unit (0) | pmol/L |
From (1) | 1 |
To (1) | 1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The stated expanded uncertainty value corresponds to the best measurement capability. |
Survey Organisation | IFCC RELA Surveys |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | INSTAND e. V. |
Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
Organization Website | http://www.instand-ev.de |
Organization Country ISO Code | DE |
Organization Country Name | Germany |
Analyte Category Name | Non-peptide hormones |