JCTLM Database: higher-order reference materials, methods and services
Details : C11RMS2 - total thyroxine (TT4) in Blood serum, Blood plasma, High purity material, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C11RMS2 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | total thyroxine (TT4) |
| IUPAC IFCC Number | total thyroxine |
| Applicable Matrices | lyophilized, fresh or frozen human serum and plasma; or calibration materials, and control specimens |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | ID-LC/MS/MS |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | LGC ID-LC/MS/MS Reference procedure for thyroxine in serum, Org-27 |
| JCTLM Database Identifier Number | CYC_II_L_I_RMP_25 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount-of-substance concentration |
| From (0) | 67.4 |
| To (0) | 234.9 |
| Unit (0) | nmol/L |
| From (1) | 1.6 |
| To (1) | 1.6 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded
uncertainty is relative uncertainty |
| Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 51. |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | Maccura |
| Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
| Organization Website | https://www.maccura.com/en/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Non-peptide hormones |