Details : C11RMS31 - estriol (non conjugated) in High purity material, Calibration solution, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC11RMS31
Reference Laboratory StatusHospital of University
Analyte Nameestriol (non conjugated)
IUPAC IFCC NumberCAS50-27-1
Traceable to SIYes
Principle of Analytical ProcedureID-LC-MS/MS
Reference Method Listed in JCTLMNo
Name of JCTLM Listed Reference Method ProcedureDGKC reference method for serum estriol
JCTLM Database Identifier NumberNRMeth 62
Procedure Difference in Your LaboratoryYes, The validation report please see 7.4.4.6.1
QuantityAmount-of-substance concentration
From (0)0.3
To (0)173.6
Unit (0)nmol/L
From (1)2.4
To (1)10
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from without dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty depends on the matrix of the sample
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 65.
Additional Information 1No. 65 is the number of our lab participation in RELA - IFCC External Quality assessment scheme for Reference Laboratories.
RemarksIncluding RELA - IFCC External Quality assessment scheme for Reference Laboratories, and PT measurement provided by New York State Department of Health Wadsworth Center and lab assigned value for Guangdong Center for Clinical Laboratories, Sichuan Maker Biotechnology CO. LTD, Shenzhen Mindray Bio-medical Electronics CO. LTD.
Recommended for Publication in the JCTLM ListYes
Validation Report for Improved MethodMethod_Validation_Report_C11RMS31.pdf
Organization AcronymGPHCM
Organization NameGuangdong Provincial Hospital of Chinese Medicine
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameNon-peptide hormones