JCTLM Database: higher-order reference materials, methods and services
Details : C11RMS32 - creatinine in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C11RMS32 |
Reference Laboratory Status | Calibration Laboratory |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID/LC/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | Isotope dilution mass spectrometry methods for creatinine in blood serum;NIST ID-LC/MS method for serum creatinine |
JCTLM Database Identifier Number | C4RMP1 |
Procedure Difference in Your Laboratory | No |
Quantity | Mass concentration |
From (0) | 0.5 |
To (0) | 5.5 |
Unit (0) | mg/dL |
From (1) | 0.03 |
To (1) | 0.25 |
Unit (1) | mg/dL |
Level of Confidence (%) | 95% |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | (b) Accredited (ISO 17025+15195) |
Survey Organisation | RELA |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results | The code of this laboratory in RELA - IFCC EQAS is 19.
Other comparison results published in Certificate of Analysis of Certified Reference Material JCCRM521-12, 4 September 2014 |
Comments of Review Teams | Range of concentration and uncertainty differs to accreditation schedule, but may be accepted as the range of concentration contracted and the range of uncertainty expanded. |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | ReCCS |
Organization Name | Reference Material Institute for Clinical Chemistry Standards |
Organization Website | http://www.reccs.or.jp/ |
Organization Country ISO Code | JP |
Organization Country Name | Japan |
Analyte Category Name | Metabolites and substrates |