JCTLM Database: higher-order reference materials, methods and services
Details : C12 RMP4R - CDC ID LC-MS/MS reference measurement procedure for total 17beta-estradiol
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C12 RMP4R |
Analyte Name | 17β-estradiol (total) |
IUPAC IFCC Number | CAS50-28-2 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | CDC ID LC-MS/MS reference measurement procedure for total 17beta-estradiol |
Organization that developed Validated Method | CDC (Centers for Control and Prevention) |
Measurement Techniques Used | ID LC-MS/MS |
Comments 1 | Calibrated with NMIJ- CRM 6004a |
Quantity | Amount-of-substance concentration |
From (0) | 2.61 |
To (0) | 3670 |
Unit (0) | pmol/L |
From (1) | 4 |
To (1) | 4.8 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Evaluation of an isotope dilution HPLC tandem mass spectrometry candidate reference measurement procedure for total 17-β estradiol in human serum, Botelho J.C. et al., Analytical Chemistry, 2016, 88(22), 11123-11129
|
Analyte Category Name | Non-peptide hormones |
Measurement principle/technique | Isotope dilution mass spectrometry |