Details : C12 RMS2 - HbA1c in Calibration solution, Whole blood
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C12 RMS2 |
Reference Laboratory Status | EQAS Organizer |
Analyte Name | HbA1c |
IUPAC IFCC Number | QU60433 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Liquid chromatography mass spectrometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | IFCC method for HbA1c (HPLC / MS-ESI) |
JCTLM Database Identifier Number | C1RMP_P16_MS |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance fraction of [HbA1c/(HbA1c+HbA0)] expressed in mmol/mol |
From (0) | 32.8 |
To (0) | 100.9 |
Unit (0) | mmol/mol |
From (1) | 1.5 |
To (1) | 1.5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded
uncertainty is a
relative uncertainty. |
Survey Organisation | IFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by
the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry
and Laboratory Medicine (DGKL), they are announced on the DGKL-RfB –
website |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | SCCL |
Organization Name | Shanghai Center for Clinical Laboratory |
Organization Website | https://www.sccl.org.cn/EnMain |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Proteins |