Details : C12RMS1 - creatinine in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC12RMS1
Reference Laboratory StatusEQAS Organizer
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Traceable to SIYes
Principle of Analytical ProcedureLiquid chromatography mass spectrometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureNIM reference method for serum creatinine
JCTLM Database Identifier NumberC7RMP26
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)58.8
To (0)485.9
Unit (0)µmol/L
From (1)1.7
To (1)1.7
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL), they are announced on the DGKL-RfB – website
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Recommended for Publication in the JCTLM ListYes
Organization AcronymSCCL
Organization NameShanghai Center for Clinical Laboratory
Organization Websitehttps://www.sccl.org.cn/EnMain
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates