Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C12RMS14 - glucose in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC12RMS14
Reference Laboratory StatusManufacturer
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Traceable to SIYes
Principle of Analytical Procedurespectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureCDC Hexokinase reference method for glucose
JCTLM Database Identifier NumberNRMeth 34
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)3.88
To (0)12
Unit (0)mmol/L
From (1)3
To (1)1.1
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range of concentration does not involve the dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Additional Information 1The participation of RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine covers the reference measurement procedures for ALT,AST,CK,LDH, GGT,AMY, ALP, K, Mg, Ca,Glu and T-Bil
Recommended for Publication in the JCTLM ListYes
Organization AcronymNingbo MedicalSystem Biotechnology Co., Ltd.
Organization NameReference Laboratory, Ningbo MedicalSystem Biotechnology Co., Ltd.
Organization Websitehttp://www.nb-medicalsystem.com/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates