JCTLM Database: higher-order reference materials, methods and services
Details : C12RMS3 - uric acid in Blood serum, Blood plasma, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C12RMS3 |
Reference Laboratory Status | Manufacturer |
Analyte Name | uric acid |
IUPAC IFCC Number | CAS69-93-2 |
Applicable Matrices | lyophilized, fresh or frozen human serum and plasma, or calibration materials, control specimens |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | NIM ID-LC/MS/MS reference method for serum Uric Acid |
JCTLM Database Identifier Number | C7RMP27 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 123.3 |
To (0) | 590.6 |
Unit (0) | µmol/L |
From (1) | 1.4 |
To (1) | 1.4 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded
uncertainty is relative uncertainty |
Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |