JCTLM Database: higher-order reference materials, methods and services

Details : C13RMP4 - HSA ID/LC/MS/MS reference measurement procedure for glucose in human serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC13RMP4
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman serum
NameHSA ID/LC/MS/MS reference measurement procedure for glucose in human serum
Organization that developed Validated MethodHealth Sciences Authority (Singapore)
Measurement Techniques UsedID/LC/MS/MS
QuantityAmount-of-substance concentration
From (0)1.5
To (0)20
Unit (0)mmol/l
From (1)0.028
To (1)0.25
Unit (1)mmol/l
Level of Confidence (%)95%
Peer Reviewed PublicationHigh accuracy analysis of glucose in human serum by isotope dilution liquid chromatography-tandem mass spectrometry, Chen Y. et al., Clinica Chimica Acta, 2012, 413(7-8), 808-813
Analyte Category NameMetabolites and substrates
Measurement principle/techniqueIsotope dilution mass spectrometry