Details : C13RMP5 - HSA ID/GC/MS reference measurement procedure for free glycerol in human serum
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C13RMP5 |
| Analyte Name | free glycerol |
| IUPAC IFCC Number | free glycerol |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | human serum |
| Name | HSA ID/GC/MS reference measurement procedure for free glycerol in human serum |
| Organization that developed Validated Method | Health Sciences Authority (Singapore) |
| Measurement Techniques Used | ID/GC/MS |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.06 |
| To (0) | 1.6 |
| Unit (0) | mmol/l |
| From (1) | 0.0018 |
| To (1) | 0.032 |
| Unit (1) | mmol/l |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Developing a reference measurement procedure for free glycerol in human serum by two-step gas chromatography-isotope dilution mass spectrometry, Liu Q. et al., Clinical Biochemistry, 2015, 48(13-14), 897-903
|
| Analyte Category Name | Metabolites and substrates |
| Measurement principle/technique | Isotope dilution mass spectrometry |