Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C13RMP5 - HSA ID/GC/MS reference measurement procedure for free glycerol in human serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC13RMP5
Analyte Namefree glycerol
IUPAC IFCC Numberfree glycerol
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman serum
NameHSA ID/GC/MS reference measurement procedure for free glycerol in human serum
Organization that developed Validated MethodHealth Sciences Authority (Singapore)
Measurement Techniques UsedID/GC/MS
QuantityAmount-of-substance concentration
From (0)0.06
To (0)1.6
Unit (0)mmol/l
From (1)0.0018
To (1)0.032
Unit (1)mmol/l
Level of Confidence (%)95%
Peer Reviewed PublicationDeveloping a reference measurement procedure for free glycerol in human serum by two-step gas chromatography-isotope dilution mass spectrometry, Liu Q. et al., Clinical Biochemistry, 2015, 48(13-14), 897-903
Analyte Category NameMetabolites and substrates
Measurement principle/techniqueIsotope dilution mass spectrometry