Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C13RMS22 - γ-glutamyltransferase (GGT) in Calibration solution, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC13RMS22
Reference Laboratory StatusManufacturer
Analyte Nameγ-glutamyltransferase (GGT)
IUPAC IFCC NumberEC 2.3.2.2
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC reference measurement procedure (37°C) for gamma-GT
JCTLM Database Identifier NumberNRMeth 69
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)0.24
To (0)4.6
Unit (0)µkat/L
From (1)4.6
To (1)4.6
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from 0.24 µkat/L to 4.60 µkat/L without dilution of the sample. The conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L concentrations measured.
Comments on Uncertainty RangeThe expanded uncertainty calculated according to GUM
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 61
Additional Information 1The participation of RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine include CK,LDH, ALT,AST,GGT,amylase,ALP.
RemarksIncluding PT measurement provided by China National Center for Clinical Laboratories and RELA - IFCC External Quality assessment scheme for Reference Laboratories, and lab assigned value for Guangdong Center for Clinical Laboratories, Shenzhen Mindray Bio-medical Electronics CO. LTD and Beijing Aerospace General Hospital Reference Laboratory
Organization AcronymShenzhen Mindray Bio-Medical Electronics Co., Ltd
Organization NameShenzhen Mindray Bio-Medical Electronics Co., Ltd. Standardization Laboratory
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes