Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMP11 - NCCL ID LC-MS/MS reference measurement procedure for glucose

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC14RMP11
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman serum; fresh, frozen or lyophilized
NameNCCL ID LC-MS/MS reference measurement procedure for glucose
Organization that developed Validated MethodNational Center for Clinical Laboratories (NCCL)
Measurement Techniques UsedIsotope Dilution Mass Spectrometry (IDMS), Liquid Chromatography Mass Spectrometry (LCMS)
QuantityAmount-of-substance fraction
From (0)1.5
To (0)25
Unit (0)mmol/L
From (1)0.5
To (1)1.5
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationDetermination of serum glucose by isotope dilution liquid chromatography-tandem mass spectrometry: a candidate reference measurement procedure, Zhang T, et al., Analytical Bioanalytical Chemistry, 2016, 408(26), 7403-7411
Additional CommentsNo
Hyperlink to Final Report of extent of Equivalence Studies amongIFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine (RELA), lab code 18, Results year 2012 and 2014
Analyte Category NameMetabolites and substrates
Non-CCQM Key Comparison ReportAgreement between the ID LC-MS/MS and the ID GC-MS reference measurement procedures for Glucose, Electronic Supplement Material of Analytical Bioanalytical Chemistry 2016 publication.
Non-CCQM Key Comparison Report Hyperlinkhttps://static-content.springer.com/esm/art%3A10.1007%2Fs00216-016-9817-0/MediaObjects/216_2016_9817_MOESM1_ESM.pdf
Measurement principle/techniqueIsotope dilution mass spectrometry
Measurement principle/techniqueLiquid chromatography mass spectrometry