Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMP5 - LC-MS/MS based candidate reference method for the quantification of carbamazepine in human serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC14RMP5
Analyte Namecarbamazepine
IUPAC IFCC NumberCAS298-46-4
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfresh or frozen human serum
NameLC-MS/MS based candidate reference method for the quantification of carbamazepine in human serum
Organization that developed Validated MethodRoche Diagnostics GmbH in collaboration with the Labormedizinisches Zentrum Dr Risch Anstalt
Measurement Techniques UsedLiquid Chromatography Mass Spectrometry (LC-MS/MS)
Comments 1as primary calibration material an ISO certified reference material was used
Comments 2carbamazepine-10,11-epoxide and 10-hydroxy-10,11-dihydrocarbamazepine were additionally monitored
QuantityMass concentration
From (0)0.1
To (0)22
Unit (0)µg/ml
Level of Confidence (%)95%
Peer Reviewed PublicationAn LC-MS/MS based candidate reference method for the quantification of carbamazepine in human serum, Taibon J. et al, Clinica Chimica Acta, 2017, 475, 35-40
Hyperlink to Final Report of extent of Equivalence Studies amongThe reference measurement method, C14 RMP5, for the quantification of carbamazepine in human serum was developed and validated by Roche Diagnostics GmbH in collaboration with the Labormedizinisches Zentrum Dr Risch Anstalt.
Analyte Category NameDrugs
Non-CCQM Key Comparison ReportSee reference cited above for comparability assessment study
Non-CCQM Key Comparison Report Hyperlink
Measurement principle/techniqueLiquid chromatography mass spectrometry