Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMP6 - NCCL ICP-MS reference measurement procedure for serum calcium

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC14RMP6
Analyte Namecalcium
IUPAC IFCC NumberCAS7440-70-2
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman fresh, frozen or lyophilized human serum
NameNCCL ICP-MS reference measurement procedure for serum calcium
Organization that developed Validated MethodNational Certer for Clinical Labortories (NCCL)
Measurement Techniques UsedInductively Coupled Plasma Mass Spectrometry (ICP-MS)
QuantityAmount-of-substance concentration
From (0)1.75
To (0)3.53
Unit (0)mmol/L
From (1)0.5
To (1)1
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationA candidate reference method for serum calcium measurement by inductively coupled plasma mass spectrometry, Yan Y. et al., Clin. Chim. Acta., 2016, 461, 141-145
Comments to be published via JCTLM DBThe expanded uncertainty is relative
Analyte Category NameElectrolytes
Non-CCQM Key Comparison ReportIFCC External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (IFCC EQAS RELA), Results 2012, 2013, 2016, Labcode=18
Non-CCQM Key Comparison Report Hyperlinkhttp://www.dgkl-rfb.de:81/index.shtml
Measurement principle/techniqueInductively coupled plasma mass spectrometry