JCTLM Database: higher-order reference materials, methods and services
Details : C14RMP7 - NCCL ICP-MS reference measurement procedure for serum potassium
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C14RMP7 |
| Analyte Name | potassium |
| IUPAC IFCC Number | CAS7440-09-7 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | human fresh, frozen or lyophilized human serum |
| Name | NCCL ICP-MS reference measurement procedure for serum potassium |
| Organization that developed Validated Method | National Certer for Clinical Labortories (NCCL) |
| Measurement Techniques Used | Inductively Coupled Plasma Mass Spectrometry (ICP-MS) |
| Quantity | Amount-of-substance concentration |
| From (0) | 1.57 |
| To (0) | 6.27 |
| Unit (0) | mmol/L |
| From (1) | 0.5 |
| To (1) | 1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | A candidate reference method for serum potassium measurement by inductively coupled plasma mass spectrometry, Yan Y. et al., Clin. Chem. Lab. Med., 2017, 55(10), 1517-1522
|
| Comments to be published via JCTLM DB | The expanded uncertainty is relative |
| Analyte Category Name | Electrolytes |
| Non-CCQM Key Comparison Report | IFCC External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (IFCC EQAS RELA), Results 2011, 2014, 2016, Labcode=18 |
| Non-CCQM Key Comparison Report Hyperlink | http://www.dgkl-rfb.de:81/index.shtml |
| Measurement principle/technique | Inductively coupled plasma mass spectrometry |