JCTLM Database: higher-order reference materials, methods and services
Details : C14RMP7 - NCCL ICP-MS reference measurement procedure for serum potassium
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C14RMP7 |
Analyte Name | potassium |
IUPAC IFCC Number | CAS7440-09-7 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | human fresh, frozen or lyophilized human serum |
Name | NCCL ICP-MS reference measurement procedure for serum potassium |
Organization that developed Validated Method | National Certer for Clinical Labortories (NCCL) |
Measurement Techniques Used | Inductively Coupled Plasma Mass Spectrometry (ICP-MS) |
Quantity | Amount-of-substance concentration |
From (0) | 1.57 |
To (0) | 6.27 |
Unit (0) | mmol/L |
From (1) | 0.5 |
To (1) | 1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | A candidate reference method for serum potassium measurement by inductively coupled plasma mass spectrometry, Yan Y. et al., Clin. Chem. Lab. Med., 2017, 55(10), 1517-1522
|
Comments to be published via JCTLM DB | The expanded uncertainty is relative |
Analyte Category Name | Electrolytes |
Non-CCQM Key Comparison Report | IFCC External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (IFCC EQAS RELA), Results 2011, 2014, 2016, Labcode=18 |
Non-CCQM Key Comparison Report Hyperlink | http://www.dgkl-rfb.de:81/index.shtml |
Measurement principle/technique | Inductively coupled plasma mass spectrometry |