Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMP9 - NCCL ICP-MS reference measurement procedure for serum sodium

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC14RMP9
Analyte Namesodium
IUPAC IFCC NumberCAS7440-23-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman fresh, frozen or lyophilized human serum
NameNCCL ICP-MS reference measurement procedure for serum sodium
Organization that developed Validated MethodNational Certer for Clinical Labortories (NCCL)
Measurement Techniques UsedInductively Coupled Plasma Mass Spectrometry (ICP-MS)
QuantityAmount-of-substance concentration
From (0)105
To (0)165
Unit (0)mmol/L
From (1)0.5
To (1)1
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationMeasurement of serum sodium and magnesium by inductively coupled plasma mass spectrometry with aluminum as internal standard, Yan Y. et al., Clin. Lab., 2016, 62, 719-725.
Comments to be published via JCTLM DBThe expanded uncertainty is relative
Analyte Category NameElectrolytes
Non-CCQM Key Comparison ReportIFCC External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (IFCC EQAS RELA), Results 2007, 2012, 2014, Labcode=18
Non-CCQM Key Comparison Report Hyperlinkhttp://www.dgkl-rfb.de:81/index.shtml
Measurement principle/techniqueInductively coupled plasma mass spectrometry