Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMS10 - glycated hemoglobin (HbA1c) in Whole blood

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC14RMS10
Reference Laboratory StatusManufacturer
Analyte Nameglycated hemoglobin (HbA1c)
IUPAC IFCC NumberQU60433
Applicable Matriceswhole blood
Traceable to SIYes
Principle of Analytical ProcedureLC/MS/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC method for HbA1c (HPLC / MS-ESI)
JCTLM Database Identifier NumberC1RMP_P16_MS
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)25
To (0)150
Unit (0)mmol/mol
From (1)1.8
To (1)1.8
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is relative uncertainty
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 51.
Organization AcronymMaccura
Organization NameMaccura Biotechnology Co., Ltd. Reference System Dept.
Organization Websitehttps://www.maccura.com/en/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameProteins