JCTLM Database: higher-order reference materials, methods and services
Details : C14RMS11 - alkaline phosphatase (ALP) in Blood serum, Blood plasma
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C14RMS11 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | alkaline phosphatase (ALP) |
| IUPAC IFCC Number | EC 3.1.3.1 |
| Applicable Matrices | human serum, or plasma; lyophilized, fresh or frozen |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | IFCC reference
measurement
procedure (37
°C) for Alkaline Phosphatase |
| JCTLM Database Identifier Number | C7RMP2R |
| Procedure Difference in Your Laboratory | No |
| Quantity | Catalytic
activity
concentration |
| From (0) | 1.07 |
| To (0) | 9.97 |
| Unit (0) | µkat/l |
| From (1) | 2.5 |
| To (1) | 2.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded
uncertainty is relative uncertainty |
| Reference Procedure Status Acceptable | YES |
| Laboratory Status | Accredited (ISO 17025+15195) |
| Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 51. |
| Participation Status in Intercomparison Surveys acceptable | YES |
| Organization Acronym | Maccura |
| Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
| Organization Website | https://www.maccura.com/en/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Enzymes |