JCTLM Database: higher-order reference materials, methods and services
Details : C14RMS15 - uric acid in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C14RMS15 |
Reference Laboratory Status | EQAS Organizer |
Analyte Name | uric acid |
IUPAC IFCC Number | CAS69-93-2 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Uric acid is determined as the area ratio of uric acid and internal standard peak by isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS) |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | NIM reference method for serum uric acid |
JCTLM Database Identifier Number | C7RMP27 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 93.5 |
To (0) | 626 |
Unit (0) | µmol/L |
From (1) | 2.4 |
To (1) | 2.4 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded
uncertainty is a
relative uncertainty. |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | b |
Survey Organisation | IFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by
the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry
and Laboratory Medicine (DGKL), they are announced on the DGKL-RfB –website |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | The laboratory participated RELA ring trials each year from 2013. The results of 2016 RELA were just at the edge of the limits of equivalence and the results of 2017 RELA (which were not published) were outside. The additional information SCCL provided indicates the corrective action is effective. The laboratory got good results both in 2018 RELA and 2019 External Quality Assessment for Reference Laboratories( in the area of China, provider is NCCL) |
Organization Acronym | SCCL |
Organization Name | Shanghai Center for Clinical Laboratory |
Organization Website | https://www.sccl.org.cn/EnMain |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |