Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMS17 - cholesterol in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC14RMS17
Reference Laboratory StatusEQAS Organizer
Analyte Namecholesterol
IUPAC IFCC NumberCAS57-88-5
Traceable to SIYes
Principle of Analytical ProcedureCholesterol is determined as the area ratio of cholesterol oxidation products and internal standard peak by isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS)
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureNCCL ID-LC/MS/MS reference procedure for serum cholesterol
JCTLM Database Identifier NumberC10RMP9
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)2
To (0)8.1
Unit (0)mmol/L
From (1)2.4
To (1)2.4
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Reference Procedure Status AcceptableYes
Laboratory Statusb
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL), they are announced on the DGKL-RfB –website
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsRecommend for listing. No Additional comments.
Organization AcronymSCCL
Organization NameShanghai Center for Clinical Laboratory
Organization Websitehttps://www.sccl.org.cn/EnMain
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates