Details : C14RMS2 - testosterone in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C14RMS2 |
Reference Laboratory Status | Manufacturer |
Analyte Name | testosterone |
IUPAC IFCC Number | CAS58-22-0 |
Applicable Matrices | lyophilized, fresh, or
frozen human serum |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC definitive method for
serum total testosterone |
JCTLM Database Identifier Number | C8RMP6 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 0.5 |
To (0) | 70 |
Unit (0) | nmol/L |
From (1) | 1.9 |
To (1) | 1.9 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded
uncertainty is relative uncertainty |
Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 51. |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Non-peptide hormones |