Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C14RMS28 - total protein in High purity material, Calibration solution, Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC14RMS28
Reference Laboratory StatusHospital of University
Analyte Nametotal protein
IUPAC IFCC Numbertotal protein
Traceable to SIYes
Principle of Analytical ProcedureBiuret Colorimetric assay
Reference Method Listed in JCTLMNo
Name of JCTLM Listed Reference Method ProcedureNo
JCTLM Database Identifier NumberNRMeth 30
Procedure Difference in Your Laboratoryyes. The validation report please see 6.4.4.6appendix 5
QuantityAmount of concentration
From (0)19
To (0)120
Unit (0)g/L
From (1)2.9
To (1)2.9
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from without dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty depends on the matrix of the sample
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 65.
Additional Information 1No. 65 is the number of our lab participation in RELA - IFCC External Quality assessment scheme for Reference Laboratories
RemarksIncluding PT measurement provided by China National Center for Clinical Laboratories and RELA - IFCC External Quality assessment scheme for Reference Laboratories, and lab assigned value for Guangdong Center for Clinical Laboratories, chongqing new health Biotechnology CO. LTD, Shenzhen Mindray Bio-medical Electronics CO. LTD, China National Accreditation Service for Conformity Assessment.
Comments for JCTLM DB PublicationThe primary calibrator used is a bovine albumin material.
Organization AcronymGPHCM
Organization NameGuangdong Provincial Hospital of Chinese Medicine
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameProteins