JCTLM Database: higher-order reference materials, methods and services
Details : C14RMS28 - total protein in High purity material, Calibration solution, Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C14RMS28 |
Reference Laboratory Status | Hospital of University |
Analyte Name | total protein |
IUPAC IFCC Number | total protein |
Traceable to SI | Yes |
Principle of Analytical Procedure | Biuret Colorimetric assay |
Reference Method Listed in JCTLM | No |
Name of JCTLM Listed Reference Method Procedure | No |
JCTLM Database Identifier Number | NRMeth 30 |
Procedure Difference in Your Laboratory | yes. The validation report please see 6.4.4.6appendix 5 |
Quantity | Amount of concentration |
From (0) | 19 |
To (0) | 120 |
Unit (0) | g/L |
From (1) | 2.9 |
To (1) | 2.9 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The measurement range spans from without dilution of the sample. |
Comments on Uncertainty Range | The expanded
uncertainty depends on
the matrix of the sample |
Survey Organisation | IFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by
the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry
and Laboratory Medicine (DGKL). |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 65. |
Additional Information 1 | No. 65 is the number of our lab participation in RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories |
Remarks | Including PT measurement provided by China National Center for Clinical Laboratories and RELA - IFCC External Quality assessment scheme for Reference Laboratories, and lab assigned value for Guangdong Center for Clinical Laboratories, chongqing new health Biotechnology CO. LTD, Shenzhen Mindray Bio-medical Electronics CO. LTD, China National Accreditation Service for Conformity Assessment. |
Comments for JCTLM DB Publication | The primary calibrator used is a bovine albumin material. |
Organization Acronym | GPHCM |
Organization Name | Guangdong Provincial Hospital of Chinese Medicine |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Proteins |