Details : C14RMS9 - creatinine in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC14RMS9
Reference Laboratory StatusManufacturer
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Applicable Matriceslyophilized, fresh, or frozen human serum
Traceable to SIYes
Principle of Analytical ProcedureID-LC/MS/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureNIM reference method for serum creatinine
JCTLM Database Identifier NumberC7RMP26
Procedure Difference in Your LaboratoryYes
QuantityAmount of substance concentration
From (0)35
To (0)2000
Unit (0)µmol/L
From (1)1.7
To (1)1.7
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is relative uncertainty
Reference Procedure Status AcceptableYes
Laboratory Statusb
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 51.
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsThe claimed range (35 - 2000 µmol/L, which is about 3.95 – 226 µg/g) is too wide. Additional evidence would be needed to support the very high concentration range.
Validation Report for Improved MethodMethod_Validation_Report_C14RMS9.pdf
Organization AcronymMaccura
Organization NameMaccura Biotechnology Co., Ltd. Reference System Dept.
Organization Websitehttps://www.maccura.com/en/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates