JCTLM Database: higher-order reference materials, methods and services
Details : C15RMS1 - haemoglobin (total) in Calibration solution, Whole blood
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C15RMS1 |
Reference Laboratory Status | Medical School Hannover |
Analyte Name | haemoglobin (total) |
IUPAC IFCC Number | CAS9008-02-0 |
Applicable Matrices | lyophilized, fresh, or frozen blood |
Traceable to SI | YES |
Principle of Analytical Procedure | Spectrophotometry (absorption of visible light applied in a procedure for measuring the amount-of-substance concentration of total haemoglobin in a liquid solution) |
Reference Method Listed in JCTLM | Not yet,
we are applying
for that as well. |
Name of JCTLM Listed Reference Method Procedure | RfB-Reference Method for Total Haemoglobin (Submitted for 2020 review) |
JCTLM Database Identifier Number | C17RMP1 |
Procedure Difference in Your Laboratory | Yes.
Validation described in own publication |
Quantity | Mass concentration |
From (0) | 20 |
To (0) | 300 |
Unit (0) | g/L |
From (1) | 1.1 |
To (1) | 1.1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | Amount-of-substance concentration range: from 1.24 mmol/L to 18.62 mmol/L |
Comments on Uncertainty Range | 1.1 % is the lowest reportable uncertainty.
(Best measurement capability, CMC) |
Laboratory Status | Accredited |
Survey Organisation | JCTLM, IFCC, RfB |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 3. |
Additional Information 1 | Results of excellent agreement to three calibration laboratories and NMI (PTB) shown in own publication |
Participation Status in Intercomparison Surveys acceptable | Y |
Remarks | The accreditation certificate (annex) refers to measurement conditions/procedure to a consensus optimization of the standard DIN 58931:2010-08. The certificate will be changed to our own publication, which is quite recently published. |
Comments of Review Teams | No objection to the service being listed on the database, would the method be accepted |
Organization Acronym | RfB |
Organization Name | Reference Institute for Bioanalytics, Calibration Laboratory II |
Organization Website | https://rfb.bio/cgi/switchLang?lang=en |
Organization Country ISO Code | DE |
Organization Country Name | Germany |
Analyte Category Name | Proteins |