JCTLM Database: higher-order reference materials, methods and services
Details : C15RMS10R - homocysteine in Blood serum
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C15RMS10R |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | homocysteine |
| IUPAC IFCC Number | CAS454-29-5 |
| Applicable Matrices | fresh, frozen or lyophilized human serum |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Isotope dilution liquid chromatography mass spectrometry (ID-LC-MS/MS) |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | NIST LC/MS/MS reference
method for serum homocysteine |
| JCTLM Database Identifier Number | NRMeth 89 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount of substance concentration |
| From (0) | 4.6 |
| To (0) | 146 |
| Unit (0) | µmol/L |
| From (1) | 1.6 |
| To (1) | 1.6 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Concentration Range | The measurement range of concentration does not involve the dilution of the sample. |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
| Organization Acronym | Ningbo MedicalSystem Biotechnology Co., Ltd. |
| Organization Name | Reference Laboratory, Ningbo MedicalSystem Biotechnology Co., Ltd. |
| Organization Website | http://www.nb-medicalsystem.com/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Metabolites and substrates |