Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C15RMS4 - alpha-amylase in Calibration solution, Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC15RMS4
Reference Laboratory StatusManufacturer
Analyte Namealpha-amylase
IUPAC IFCC NumberEC 3.2.1.1
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC reference measurement procedure (37 °C) for amylase
JCTLM Database Identifier NumberNRMeth 70
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)0.725
To (0)9.752
Unit (0)µkat/l
From (1)2.7
To (1)1.7
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty
Reference Procedure Status AcceptableYES
Laboratory Status(b)
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 63.
Participation Status in Intercomparison Surveys acceptableYES
Comments of Review TeamsMaterial: IFCC PRMP not for high purity mat. not accredited for high purity mat.
Organization AcronymLEADMAN
Organization NameReference Laboratory of Beijing Leadman Biochemistry Co., Ltd
Organization Websitehttp://www.leadmanbio.com
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes