JCTLM Database: higher-order reference materials, methods and services
Details : C15RMS9 - HbA1c in Whole blood, Calibration solution, Other
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C15RMS9 |
| Reference Laboratory Status | EQAS Organizer |
| Analyte Name | HbA1c |
| IUPAC IFCC Number | QU60433 |
| Applicable Matrices | Whole blood, frozen or lyophilized haemolysates |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Hemoglobin samples treated with endoproteinase Glu-C followed by high performance liquid chromatography (HPLC) and electrospray ionisation mass spectrometry (ESI-MS) |
| Reference Method Listed in JCTLM | Yes |
| JCTLM Database Identifier Number | C1RMP_P16_MS |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount-of-substance fraction |
| From (0) | 29 |
| To (0) | 150 |
| Unit (0) | mmol/mol |
| From (1) | 2 |
| To (1) | 2 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The stated expanded uncertainty value corresponds to the best measurement capability. |
| Survey Organisation | RELA - IFCC External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
| Reference to Published Comparison Results | RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine.
The code of this laboratory in RELA - IFCC EQAS is 1. |
| Additional Information 1 | RELA results 2020 published without, but 2021 and 2022 results with disclosure of laboratory identity. The code of this laboratory in RELA is 1 |
| Remarks | Approved Reference Laboratory of the IFCC HbA1c network; Laboratory provides traceable target values for EQAS regularly. Laboratory contributes to the EurA1c study as reference laboratory (target value determination) since 2021. |
| Comments for JCTLM DB Publication | Approved Reference Laboratory of the IFCC HbA1c network. Laboratory provides traceable target values for EQAS regularly. Laboratory contributes to the EurA1c study as reference laboratory (target value determination) since 2021. |
| Organization Acronym | SPMD-RfB |
| Organization Name | Reference Institute for Bioanalysis, Calibration Laboratory I |
| Organization Website | https://www.rfb.bio |
| Organization Country ISO Code | DE |
| Organization Country Name | Germany |
| Analyte Category Name | Proteins |