JCTLM Database: higher-order reference materials, methods and services
Details : C16RMP13R - GPHCM Reference measurement method for serum estriol
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C16RMP13R |
| Analyte Name | estriol (non conjugated) |
| IUPAC IFCC Number | CAS50-27-1 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | Traceable to SI |
| Applicable Matrices | human serum; fresh, frozen or lyophilized |
| Name | GPHCM Reference measurement method for serum estriol |
| Organization that developed Validated Method | Guangdong Provincial Hospital of Chinese Medicine (GPHCM) |
| Measurement Techniques Used | Isotope dilution liquid chromatography mass spectrometry |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.35 |
| To (0) | 173.38 |
| Unit (0) | nmol/L |
| From (1) | 2.4 |
| To (1) | 10 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Measurement of human serum unconjugated estriol without derivatization using liquid chromatography-tandem mass spectrometry candidate reference method and compared with two immunoassays, Huang X. et al,
Analytical and Bioanalytical Chemistry, 2018, 401(24), 6257-6267
|
| Hyperlink to Final Report of extent of Equivalence Studies among | C16RMP13R_Extent-of-equivalence_study_Report-estriol.pdf |
| Analyte Category Name | Non-peptide hormones |
| Non-CCQM Key Comparison Report | C16RMP13R Inter-laboratory Comparison Report |
| Non-CCQM Key Comparison Report Hyperlink | https://www.bipm.org/jctlm/final_report/C16RMP13R_Inter-laboratory_Comparison_Report-estriol.pdf |
| Measurement principle/technique | Isotope dilution mass spectrometry |
| Measurement principle/technique | Liquid chromatography mass spectrometry |