Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C16RMP6 - ID LC-MS/MS based candidate reference method for the quantification of androstenedione in human serum and plasma

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC16RMP6
Analyte Nameandrostenedione
IUPAC IFCC Numberandrostenedione
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfresh or frozen human serum and plasma
NameID LC-MS/MS based candidate reference method for the quantification of androstenedione in human serum and plasma
Organization that developed Validated MethodRoche Diagnostics GmbH in collaboration with the Klinik für Kinder und Jugendliche, Universitätsklinikum Erlangen
Measurement Techniques UsedIsotope dilution mass spectrometry
QuantityMass concentration
From (0)0.05
To (0)12
Unit (0)ng/mL
From (1)0.0058
To (1)0.676
Unit (1)ng/mL
Level of Confidence (%)95 %
Peer Reviewed PublicationAn isotope dilution LC-MS/MS based candidate reference method for the quantification of androstenedione in human serum and plasma, Gradl K. et al., Clinical Mass Spectrometry, 2020, 16, 1-10
Comments to be published via JCTLM DBThe reference measurement procedure, C16 RMP6, for quantification of androstenedione in human serum and plasma was developed and validated by Roche Diagnostics GmbH in collaboration with the Klinik für Kinder und Jugendliche, Universitätsklinikum Erlangen
Analyte Category NameNon-peptide hormones
Measurement principle/techniqueIsotope dilution mass spectrometry