JCTLM Database: higher-order reference materials, methods and services
Details : C16RMP7R - ID LC-MS/MS reference measurement method for cyclosporine A in human whole blood
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C16RMP7R |
Analyte Name | cyclosporin A |
IUPAC IFCC Number | CAS59865-13-3 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | frozen human whole blood;(EDTA, CDT) |
Name | ID LC-MS/MS reference measurement method for cyclosporine A in human whole blood |
Organization that developed Validated Method | Roche Diagnostics GmbH in collaboration with Institute of Laboratory Medicine, Hospital of the Ludwig-Maximilians University, and Labormedizinisches Zentrum Dr Risch Anstalt |
Measurement Techniques Used | isotope dilution mass spectrometry |
Comments 1 | tacrolimus, sirolimus and everolimus were quantified within the same mehod |
Quantity | Mass concentration |
From (0) | 5 |
To (0) | 2100 |
Unit (0) | ng/mL |
From (1) | 5.5 |
To (1) | 7.2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | An isotope dilution LC-MS/MS based candidate reference method for the quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human whole blood, Taibon J. et al.,
|
Comments to be published via JCTLM DB | This reference measurement method was developed and validated by Roche Diagnostics GmbH in collaboration with the Institute of Laboratory Medicine, Hospital of the Ludwig-Maximilians University, and Labormedizinisches Zentrum |
Analyte Category Name | Drugs |
Measurement principle/technique | Isotope dilution mass spectrometry |