Details : C16RMP9R - ID LC-MS/MS reference measurement method for sirolimus in human whole blood

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC16RMP9R
Analyte Namesirolimus
IUPAC IFCC NumberCAS53123-88-9
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfrozen human whole blood;(EDTA, CDT)
NameID LC-MS/MS reference measurement method for sirolimus in human whole blood
Organization that developed Validated MethodRoche Diagnostics GmbH in collaboration with Institute of Laboratory Medicine, Hospital of the Ludwig-Maximilians University, and Labormedizinisches Zentrum Dr Risch Anstalt
Measurement Techniques Usedisotope dilution mass spectrometry
Comments 1cyclosporine A, everolimus and tacrolimus were quantified within the same mehod
QuantityMass concentration
From (0)0.25
To (0)50
Unit (0)ng/mL
From (1)5.8
To (1)9
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationAn isotope dilution LC-MS/MS based candidate reference method for the quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human whole blood, Taibon J. et al.,
Comments to be published via JCTLM DBThis reference measurement method was developed and validated by Roche Diagnostics GmbH in collaboration with the Institute of Laboratory Medicine, Hospital of the Ludwig-Maximilians University, and Labormedizinisches Zentrum
Analyte Category NameDrugs
Measurement principle/techniqueIsotope dilution mass spectrometry