JCTLM Database: higher-order reference materials, methods and services
Details : C16RMS13 - alanine aminotransferase (ALT) in Blood serum, Blood plasma, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C16RMS13 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | alanine aminotransferase (ALT) |
| IUPAC IFCC Number | EC2.6.1.2 |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Kinetic spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | IFCC reference measurement
procedure (37 °C) for ALT |
| JCTLM Database Identifier Number | NRMeth 67 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Catalytic activity concentration |
| From (0) | 0.24 |
| To (0) | 4.3 |
| Unit (0) | µkat/l |
| From (1) | 2.1 |
| To (1) | 2.1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Concentration Range | The conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty |
| Reference Procedure Status Acceptable | yes |
| Laboratory Status | (b) |
| Survey Organisation | JCTLM(IFCC),www.dgkl-rfb.de:81 |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 138. |
| Participation Status in Intercomparison Surveys acceptable | yes |
| Organization Acronym | AutoRF |
| Organization Name | Reference laboratory of Beijing, Autobio Biochemistry Co., Ltd |
| Organization Website | http://www.autobio-bj.com.cn/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Enzymes |