Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C16RMS15 - alpha-amylase (AMY) in Blood serum, Blood plasma, Calibration solution

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC16RMS15
Reference Laboratory StatusManufacturer
Analyte Namealpha-amylase (AMY)
IUPAC IFCC NumberEC3.2.1.1
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC reference measurement procedure (37 °C) for amylase
JCTLM Database Identifier NumberNRMeth 70
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)0.57
To (0)14.73
Unit (0)µkat/l
From (1)2.6
To (1)2.6
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty
Reference Procedure Status Acceptableyes
Laboratory Status(b)
Survey OrganisationJCTLM(IFCC),www.dgkl-rfb.de:81
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 138.
Participation Status in Intercomparison Surveys acceptableyes
Organization AcronymAutoRF
Organization NameReference laboratory of Beijing, Autobio Biochemistry Co., Ltd
Organization Websitehttp://www.autobio-bj.com.cn/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes