Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C16RMS20 - glucose in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC16RMS20
Reference Laboratory StatusManufacturer
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Traceable to SIYes
Principle of Analytical Procedurespectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureCDC Hexokinase reference method for glucose
JCTLM Database Identifier NumberNRMeth 34
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)3.98
To (0)17.71
Unit (0)mmol/L
From (1)1.6
To (1)1.6
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty
Survey OrganisationJCTLM(IFCC),www.dgkl-rfb.de:81
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 138.
Organization AcronymAutoRF
Organization NameReference laboratory of Beijing, Autobio Biochemistry Co., Ltd
Organization Websitehttp://www.autobio-bj.com.cn/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates