JCTLM Database: higher-order reference materials, methods and services
Details : C16RMS20 - glucose in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C16RMS20 |
Reference Laboratory Status | Manufacturer |
Analyte Name | glucose |
IUPAC IFCC Number | CAS50-99-7 |
Traceable to SI | Yes |
Principle of Analytical Procedure | spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC Hexokinase reference method for glucose |
JCTLM Database Identifier Number | NRMeth 34 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 3.98 |
To (0) | 17.71 |
Unit (0) | mmol/L |
From (1) | 1.6 |
To (1) | 1.6 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty |
Survey Organisation | JCTLM(IFCC),www.dgkl-rfb.de:81 |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results |
The code of this laboratory in RELA - IFCC EQAS is 138. |
Organization Acronym | AutoRF |
Organization Name | Reference laboratory of Beijing, Autobio Biochemistry Co., Ltd |
Organization Website | http://www.autobio-bj.com.cn/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |