Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C16RMS21 - urea in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC16RMS21
Reference Laboratory StatusManufacturer
Analyte Nameurea
IUPAC IFCC NumberCAS57-13-6
Traceable to SIYes
Principle of Analytical Procedurespectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureCDC reference method for urea
JCTLM Database Identifier NumberNRMeth 36
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)2.57
To (0)28.5
Unit (0)mmol/L
From (1)1.9
To (1)1.9
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty
Survey OrganisationJCTLM(IFCC),www.dgkl-rfb.de:81
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison Results The code of this laboratory in RELA - IFCC EQAS is 138.
Organization AcronymAutoRF
Organization NameReference laboratory of Beijing, Autobio Biochemistry Co., Ltd
Organization Websitehttp://www.autobio-bj.com.cn/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates