JCTLM Database: higher-order reference materials, methods and services
Details : C16RMS30R - total cholesterol in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C16RMS30R |
Reference Laboratory Status | Manufacturer |
Analyte Name | total cholesterol |
IUPAC IFCC Number | total cholesterol |
Applicable Matrices | Human serum; lyophilized, fresh, or frozen |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS/MS |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | C10RMP9 |
Procedure Difference in Your Laboratory | Yes |
Quantity | Amount-of-substance concentration |
From (0) | 1.377 |
To (0) | 9.151 |
Unit (0) | mmol/L |
From (1) | 2.5 |
To (1) | 2.5 |
Unit (1) | % |
Level of Confidence (%) | 95 % |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine.
The code of this laboratory in RELA-IFCC EQAS is 051. |
Validation Report for Improved Method | Validation report of the ID-LC-MSMS method for Total cholesterol.pdf |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |