JCTLM Database: higher-order reference materials, methods and services
Details : C17RMP1 - RfB-Reference method for total haemoglobin
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C17RMP1 |
| Analyte Name | haemoglobin (total) |
| IUPAC IFCC Number | CAS9008-02-0 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen whole blood |
| Name | RfB-Reference method for total haemoglobin |
| Organization that developed Validated Method | Reference Institute for Bioanalytics, Calibration Laboratory II |
| Measurement Techniques Used | Spectrophotometry |
| Comments 1 | The reference measurement procedure determines the concentration of total haemoglobin. |
| Quantity | Mass concentration; amount-of-substance concentration |
| From (0) | 20 |
| To (0) | 300 |
| Unit (0) | g/L |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | 1.1 % and above depending on the matrix properties |
| Peer Reviewed Publication | Total haemoglobin - a reference measuring system for improvement of standardization, Grote-Koska D. et al, Clinical Chemistry and Laboratory Medicine, 2020, 58(8), 1314-1321
|
| Comments to be published via JCTLM DB | The reference measurement procedure gives total concentration of the mass and of the amount-of-substance of haemoglobin in the sample material. |
| Analyte Category Name | Proteins |
| Non-CCQM Key Comparison Report | IFCC RELA EQA |
| Non-CCQM Key Comparison Report Hyperlink | http://www.dgkl-rfb.de:81/index.shtml |
| Measurement principle/technique | Spectrophotometry |