Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C17RMP1 - RfB-Reference method for total haemoglobin

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC17RMP1
Analyte Namehaemoglobin (total)
IUPAC IFCC NumberCAS9008-02-0
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen whole blood
NameRfB-Reference method for total haemoglobin
Organization that developed Validated MethodReference Institute for Bioanalytics, Calibration Laboratory II
Measurement Techniques UsedSpectrophotometry
Comments 1The reference measurement procedure determines the concentration of total haemoglobin.
QuantityMass concentration; amount-of-substance concentration
From (0)20
To (0)300
Unit (0)g/L
Level of Confidence (%)95%
Comments on Uncertainty Range1.1 % and above depending on the matrix properties
Peer Reviewed PublicationTotal haemoglobin - a reference measuring system for improvement of standardization, Grote-Koska D. et al, Clinical Chemistry and Laboratory Medicine, 2020, 58(8), 1314-1321
Comments to be published via JCTLM DBThe reference measurement procedure gives total concentration of the mass and of the amount-of-substance of haemoglobin in the sample material.
Analyte Category NameProteins
Non-CCQM Key Comparison ReportIFCC RELA EQA
Non-CCQM Key Comparison Report Hyperlinkhttp://www.dgkl-rfb.de:81/index.shtml
Measurement principle/techniqueSpectrophotometry