Details : C17RMP2 - HSA ID-LC-MS/MS reference measurement procedure for testosterone in human serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC17RMP2
Analyte Nametestosterone
IUPAC IFCC NumberCAS58-22-0
Method capable of Traceability of Analyte to SI or Defined ProcedureTraceable to SI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameHSA ID-LC-MS/MS reference measurement procedure for testosterone in human serum
Organization that developed Validated MethodHealth Sciences Authority (HSA), Singapore
Measurement Techniques UsedID-LC-MS/MS
QuantityAmount-of-substance concentration
From (0)0.5
To (0)40
Unit (0)nmol/L
From (1)0.03
To (1)0.8
Unit (1)nmol/L
Level of Confidence (%)95%
Peer Reviewed PublicationSimple and accurate candidate reference measurement procedure for total testosterone in human serum by one-step liquid-liquid extraction coupled with isotope dilution mass spectrometry, Chen Y. et al Analytical and Bioanalytical Chemistry, 2019, 411(28), 7519-7528
Analyte Category NameNon-peptide hormones
Measurement principle/techniqueIsotope dilution mass spectrometry
Measurement principle/techniqueLiquid chromatography mass spectrometry