JCTLM Database: higher-order reference materials, methods and services
Details : C17RMP2 - HSA ID-LC-MS/MS reference measurement procedure for testosterone in human serum
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C17RMP2 |
Analyte Name | testosterone |
IUPAC IFCC Number | CAS58-22-0 |
Method capable of Traceability of Analyte to SI or Defined Procedure | Traceable to SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | HSA ID-LC-MS/MS reference measurement procedure for testosterone in human serum |
Organization that developed Validated Method | Health Sciences Authority (HSA), Singapore |
Measurement Techniques Used | ID-LC-MS/MS |
Quantity | Amount-of-substance concentration |
From (0) | 0.5 |
To (0) | 40 |
Unit (0) | nmol/L |
From (1) | 0.03 |
To (1) | 0.8 |
Unit (1) | nmol/L |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Simple and accurate candidate reference measurement procedure for total testosterone in human serum by one-step liquid-liquid extraction coupled with isotope dilution mass spectrometry, Chen Y. et al
Analytical and Bioanalytical Chemistry, 2019, 411(28), 7519-7528
|
Analyte Category Name | Non-peptide hormones |
Measurement principle/technique | Isotope dilution mass spectrometry |
Measurement principle/technique | Liquid chromatography mass spectrometry |