Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C17RMP4 - CDC equilibrium dialysis ID LC-MS/MS reference measurement procedure for free thyroxine in human serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC17RMP4
Analyte Namefree thyroxine
IUPAC IFCC Numberfree thyroxine
Method capable of Traceability of Analyte to SI or Defined ProcedureTraceable to SI
NameCDC equilibrium dialysis ID LC-MS/MS reference measurement procedure for free thyroxine in human serum
Organization that developed Validated MethodCenters for Control and Prevention,CDC
Measurement Techniques UsedEquilibrium dialysis isotope dilution-liquid chromatography tandem mass spectrometry (ID LC-MS/MS)
Comments 1Calibrated with IRMM 468
Quantity
Amount-of-substance concentration
From (0)1.29
To (0)258
Unit (0)pmol/L
From (1)7
To (1)8.8
Unit (1)%
Level of Confidence (%)95 %
Peer Reviewed PublicationDevelopment of an equilibrium dialysis ID-UPLC-MS/MS candidate reference measurement procedure for free thyroxine in human serum, Zhang L. et al, Clinical Biochemistry 2023 116:42-51
Analyte Category NameNon-peptide hormones
Non-CCQM Key Comparison ReportSee reference cited for comparability assessment study
Measurement principle/techniqueIsotope dilution mass spectrometry