Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C17RMS16 - alkaline phosphatase (ALP) in Blood plasma, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC17RMS16
Reference Laboratory StatusManufacturer
Analyte Namealkaline phosphatase (ALP)
IUPAC IFCC NumberEC3.1.3.1
Applicable MatricesSerum or plasma; lyophilized powder, fresh or frozen
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectroscopy
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberC7RMP2R
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)1.11
To (0)9.95
Unit (0)µkat/l
From (1)2.18
To (1)2.15
Unit (1)%
Level of Confidence (%)95 %
Comments on Concentration RangeThe conversion factor for enzyme catalytic activity concentration: 1 U/L = 0.01667 µkat/L.
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at http://www.dgkl- rfb.de:81/index.shtml The code of this laboratory in RELA - IFCC EQAS is 103.
Intercomparison Results Publicly AvailableYES
Link to Comparison Resultswww.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison ResultsRELA-IFCC External quality assessment scheme for Reference Laboratories in Laboratory Medicine. The code of this laboratory in RELA-IFCC EQAS is 103.
Organization AcronymFosun Diagnostics
Organization NameFosun Diagnostics (Shanghai) Co., Ltd
Organization Websitehttps://www.fosundiagnostics.com
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes