JCTLM Database: higher-order reference materials, methods and services
Details : C17RMS17 - α-amylase (AMY) in Blood plasma, Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C17RMS17 |
Reference Laboratory Status | Manufacturer |
Analyte Name | α-amylase (AMY) |
IUPAC IFCC Number | EC3.2.1.1 |
Applicable Matrices | Serum or plasma; lyophilized powder, fresh or frozen |
Traceable to SI | Yes |
Principle of Analytical Procedure | Kinetic spectroscopy |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | NRMeth 70 |
Procedure Difference in Your Laboratory | No |
Quantity | Catalytic activity concentration |
From (0) | 0.62 |
To (0) | 11.23 |
Unit (0) | µkat/l |
From (1) | 2.4 |
To (1) | 2.28 |
Unit (1) | % |
Level of Confidence (%) | 95 % |
Comments on Concentration Range | The conversion factor for enzyme catalytic activity concentration: 1 U/L = 0.01667 µkat/L. |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty.
|
Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in Laboratory Medicine at
http://www.dgkl-
rfb.de:81/index.shtml
The code of this
laboratory in RELA
- IFCC EQAS is 103.
|
Intercomparison Results Publicly Available | YES |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA-IFCC External quality assessment scheme for Reference Laboratories in Laboratory Medicine.
The code of this laboratory in RELA-IFCC EQAS is 103. |
Organization Acronym | Fosun Diagnostics |
Organization Name | Fosun Diagnostics (Shanghai) Co., Ltd |
Organization Website | https://www.fosundiagnostics.com |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Enzymes |